    10.10: Checking Count Quality

    Technical repeats are repeated measurements of the same sample. Loading two chambers can help reveal poor mixing or loading differences. They do not represent two independent cultures. Follow the counting plan and review the separate results before averaging them.

    For example, chamber averages of 49 and 51 clear cells per large square are close. Their mean is 50. With a factor of 2, that mean gives 1.0 × 10⁶ viable cells/mL. In contrast, averages of 42 and 85 deserve investigation. An average of those two numbers may look reasonable while hiding a sampling problem.

    Check for settling, clumps, bubbles, incorrect chamber filling, different areas counted, or a transcription error. If justified, prepare and load a fresh sample according to the procedure. Keep the original results and explain the repeat. Do not erase an inconvenient count or keep repeating until the desired answer appears.

    Measurement uncertainty is the doubt associated with a measurement. A count based on very few cells is less reliable than a larger count under suitable conditions. Counting more areas can reduce random sampling variation. It cannot fix a repeated dilution mistake. Use the method’s limits for repeat agreement rather than choosing a percentage after seeing the data.


    This page titled 10.10: Checking Count Quality was last modified on Wed, 21 Oct 2026 01:34:10 GMT and is shared under a CC BY-NC-SA 4.0 license and was authored, remixed, and/or curated by Emalee MacKenzie.