An automated cell counter uses an instrument to estimate cell number, often through imaging or an electrical measurement. Review the method, counted cell size range, and sample limits. For an imaging counter, inspect the stored image when possible. Clumps or debris can still confuse the analysis. Check whether the instrument already corrects for dye dilution. Some systems apply a twofold correction when an equal-volume trypan blue setting is selected. Multiplying that reported concentration by 2 again would double the result incorrectly. Record relevant settings, dilution corrections, and the original output.
Other viability assays measure different cell properties. ATP tests and metabolic assays can estimate viable populations under suitable conditions. Their signals also depend on cell activity and assay conditions. A lower signal may mean fewer viable cells, lower activity per cell, or measurement interference. Appropriate controls help distinguish these causes. Choose the test that fits the question. A routine membrane-integrity count helps plan a seeding volume. A treatment study may need additional evidence of growth, survival, or biological function. Report the method with the result so that “90% viable” has a clear meaning.