Accuracy describes how close a measurement is to the accepted value. Precision describes how closely repeated measurements agree.
If a device repeatedly delivers 90 microliters when set to 100, its transfers may be precise but inaccurate. If repeated transfers vary widely, precision is poor. Both problems can affect an experiment.
Calibration evaluates an instrument against a reference and may support adjustment under an approved process. Verification checks whether specified performance requirements are met. A pre-use inspection is useful, but it is not a full calibration.
A pipette check can include the instrument identity, approved service status, physical condition, tip fit, and signs of leakage. A laboratory may also use a performance check based on a defined procedure. Record the actual check rather than writing “calibrated” without evidence.
Consider a dilution intended to deliver the same treatment to every well. A systematic pipette error can shift all doses. Variable delivery can create differences among wells. Careful volume transfer therefore supports both treatment comparability and sound interpretation.
"How to Calibrate Your Pipette" DLAB Scientific Inc. https://www.youtube.com/watch?v=rsgnXlRELVA Standard YouTube License. No changes made.