A centrifuge spins samples to separate components according to their behavior 
under centrifugal force. A rotor holds the tubes or other vessels during rotation.
After centrifugation, collected material may form a pellet. The liquid above it
is the supernatant. Whether the experiment needs the pellet, the supernatant, or both depends on the purpose. A pellet may be small or difficult to see.
Revolutions per minute, or rpm, describes rotational speed. Relative centrifugal force, or RCF, describes force relative to gravity and is commonly written as a multiple of g. The same rpm can produce different RCF values in rotors with different radii.
A commonly used relationship is:
RCF = 1.118 × 10⁻⁵ × r × rpm²
Here, r is the applicable rotor radius in centimeters. The rotor documentation identifies which radius should be used for the intended setting.
Worked example: A rotor with a specified radius of 10 cm operates at 2,000 rpm.
RCF = 1.118 × 10⁻⁵ × 10 × 2,000² = 447.2
The resulting force is approximately 447 × g. A different rotor radius would give a different value at 2,000 rpm.
A device that allows direct RCF entry may simplify setup, but the selected rotor information must still be correct. Time, temperature, acceleration, and braking can also affect separation. Use all the conditions specified for the sample.