    2.11: Centrifuge Loading and Care

    Before starting the centrifuge, take a moment to check how the rotor is loaded. Follow the instrument’s approved loading pattern and balance the complete loads, including tubes, caps, adapters, and liquid. Two tubes may look equally full but have different masses because their containers or contents differ. Check the loaded masses rather than relying on appearance alone.

    As you load the samples, inspect the rotor, tubes, adapters, and seals. Look for cracks, corrosion, or other damage, and confirm that each part is compatible with the instrument and intended settings. A small defect is easier to address before the run than after the rotor begins turning. If you are unsure about a part, ask the instructor or responsible supervisor before using it.

    Consider what could happen if a tube leaked or broke. Some biological samples require sealed rotors or safety cups to help contain aerosols, along with a specific procedure for opening them. Follow the laboratory’s risk assessment and response instructions. If breakage is suspected, keep the instrument closed and obtain guidance before opening it.

    Close the centrifuge as directed and start the run. Unusual vibration or noise is a reason to act: stop the run using the instrument controls and obtain assistance. Wait until the rotor has stopped completely before opening the instrument, and never bypass the lid interlock.

    Record the rotor used, force or speed, run time, temperature when relevant, and any unexpected event. These details help another person understand what happened to the samples. After use, follow the approved cleaning and maintenance procedures. If a rotor is dropped or damaged, report it and have it assessed before it is used again.