Calibration establishes a relationship between a measurement indication and reference values under specified conditions, including relevant measurement uncertainty. It may identify bias or support a correction. Adjustment changes the measuring system and is a separate activity, even when performed during the same service visit.
A calibration report may include results before and after adjustment, often called as-found and as-left results. Not every calibration requires adjustment or a pass/fail decision. A company must assess whether the uncertainty and performance are suitable for the intended measurement.
Verification provides objective evidence that specified requirements are met. A balance check using a suitable weight or a pipette performance check may be verification activities. Their adequacy depends on the procedure, acceptance criteria, reference, and intended use. Verification does not automatically replace required calibration.
Qualification documents that facilities, utilities, or equipment are suitable for their intended use. The scope may include design review and installation, operational, and performance qualification. Use qualified personnel and appropriate evidence rather than assuming that a supplier’s initial calibration proves suitability after shipment and installation.
Validation demonstrates that a process or method is suitable for its intended purpose using scientifically justified evidence. It is not proof of zero error under every imaginable condition. Equipment qualification supports process validation but does not replace it.

This page titled 14.11: Calibration Verification Qualification and Validation was last modified on Sun, 08 Nov 2026 23:43:05 GMT and is shared under a CC BY-SA 4.0 license and was authored, remixed, and/or curated by Emalee MacKenzie.