    14.12: Process Validation Across the Life Cycle

    Qualifying Equipment for Intended Use.pngInstallation qualification (IQ) checks that equipment and supporting systems are installed according to approved requirements. Evidence may include identity, model and serial number, location, materials of construction, utilities, wiring and connections, drawings, manuals, and relevant calibration or maintenance arrangements.

    Operational qualification (OQ) evaluates operation over intended operating ranges. It may challenge controls, alarms, interlocks, functions, and relevant limits. Acceptance criteria and test conditions should be justified for the equipment and use.

    Equipment performance qualification (equipment PQ) evaluates reliable performance for the intended use under representative conditions, such as defined loads or operating patterns. Its relationship to process qualification depends on the system and validation strategy.

    For a steam sterilizer, installation checks address the unit and utilities; operational checks evaluate relevant controls and operation; load-specific studies assess effective steam penetration and exposure for the intended loads. The location most difficult to sterilize must be determined through evidence, not assumed to be a particular chamber corner.

    An equipment file should include its unique identifier; manufacturer, model, and serial number; receipt and installation information; intended use; qualification evidence; relevant calibration; instructions; maintenance; training or procedure references; and a history of failures, repairs, modifications, and status decisions.

    If a measuring instrument is found outside acceptance criteria, control its use and investigate the effect on previous measurements and products. Adjustment or recalibration may be needed, followed by appropriate checks. Full revalidation is not automatically required for every calibration issue; the impact assessment determines the necessary action.