    14.2: Facilities and Equipment in Production

    A facility must support manufacture of products that meet applicable quality requirements. Not every pharmaceutical product is sterile. Nonsterile products still require appropriate contamination controls, while sterile products require additional controls suited to their processing strategy.

    Facility layout, personnel practices, material flow, cleaning, and storage should reduce contamination, cross-contamination, and mix-ups. Contamination means unwanted material is present. Cross-contamination involves transfer from another product or material. A mix-up occurs when the wrong material, product, label, or instruction is used; it is a different type of error.

    Control the status and movement of sterile and nonsterile items. Protect exposed product and critical operations. Keep accepted, quarantined, and rejected materials distinguishable through appropriate physical segregation or other adequately controlled systems. The correct approach depends on the risks and applicable requirements.

    Housekeeping and sanitation programs should keep the facility clean and control pests. Environmental monitoring should be appropriate to the product and process. Monitoring cannot replace a well-designed facility and effective contamination prevention.

    Equipment and instruments must have sufficient capacity and accuracy for the task. Establish maintenance, calibration, cleaning, operating, and malfunction procedures. Equipment status should be clear to users, and failures should be reported promptly. Evaluate whether an equipment problem could have affected previously generated data or manufactured material.

    A freezer may operate near −80 °C, and a refrigerator may have a defined refrigerated range, but actual limits must be established for the equipment and stored materials. A display reading alone does not prove that every location in a storage unit is within range. Mapping, monitoring, alarms, and calibration may be needed according to the intended use.

    Steam sterilization depends on effective saturated steam contact, temperature, exposure time, air removal, and the load. Pressure supports the required steam conditions; it is not an independent guarantee of sterilization. Common steam sterilization temperatures include 121 °C and 132 °C, but cycle settings and exposure times must be validated for the sterilizer, load, and application. Follow the approved cycle and manufacturer’s instructions.