    8.12: Johnson Noise Thermometry (JNT)

    A Johnson noise thermometer[17, 21, 22, 23] determines the absolute temperature T by measuring the mean-square voltage fluctuations \(<V^{2}>\) across a sensing resistor \(R\), governed by the Nyquist formula given above Eq.(1):

    \[ S_{V}(\omega)=4 k_{B} T R . \nonumber \]

    However, in practical measurement systems [24, 25, 26], the sensing resistor is connected to amplifiers through leads and circuitry that introduce parasitic capacitance \(C\) and inductance \(L\). As a result, the impedance becomes frequency dependent \(Z(\omega)\), and \(R\) above is replaced by \(\operatorname{Re}(Z)\) as in Eq.(2), and the "noise power" is integrated over a specific bandwidth where the ratio of the real part of the impedance to the total gain is precisely known.

    Modern precision methods, such as those developed by NIST[24, 25], often utilize a dual-path cross-correlation technique or a combining of thermal noise with a synthesized reference signal to eliminate the influence of parasitic components. By measuring the complex transmission line characteristics using an alternating current (AC) quantized voltage source or a ratio-metric bridge, researchers can determine the exact attenuation and phase shift caused by the parasitics. This allows for the "de-embedding" of the sensor's true noise from the hardware's electrical signature, ensuring the temperature remains traceable directly to the Boltzmann constant without being skewed by the physical wiring of the cryostat or probe.

    A review of Johnson noise thermometry is beyond the scope of this book and one should consult the papers cited here and references there in. JNT is particularly important with the

    recent redefinition of the kelvin, (symbol \(K\) ). As of May 2019, the National Institute of Standards and Technology (NIST) redefined the kelvin \((K)\) based on the Boltzmann constant \(\left(K_{B}\right)\) rather than the triple point of water. The magnitude is now fixed by taking the fixed numerical value of the Boltzmann constant \(K_{B}\) to be \(1.380649 \times 10^{-23}\) when expressed in the unit Joules per Kelvin \(J K^{-1}\), which is equal to \(k g m^{2} s^{-2} K^{-1}\).

    Johnson noise thermometry (JNT) is critically important following the 2019 redefinition of the kelvin because it acts as a primary, purely electronic method to realize and disseminate thermodynamic temperature directly from the fixed Boltzmann constant. With the redefinition, the kelvin is no longer based on the triple point of water, allowing temperature to be measured independent of material properties, making JNT a versatile tool for high-accuracy.