14.4: Tangential Flow Filtration

Cell Culture Textbook (Biot 77 IVC) · LibreTexts

0%

Contents

    Text

    Size
    100%
    Theme
    Typeface
    Line length
    Spacing

    Read aloud

    How

    Speed: 1×
    Light up
    Highlight

    Symbols: Mulberry Symbols, © 2018–2026 Steve Lee, CC BY-SA 4.0 (licence, the set), the site's own copies.

    From LibreTexts, under CC BY-NC-SA 4.0, as the source states it; its equations are written in TeX, which the PDF shows typeset. This page at LibreTexts.

    14.4: Tangential Flow Filtration

    Tangential flow filtration, or TFF, moves liquid along a membrane. Some liquid and material cross through it. In dead-end filtration, flow is directed toward the membrane instead. Cross flow can reduce buildup but cannot eliminate fouling. TFF can concentrate an EV-containing sample and exchange its solution. It does not by itself establish EV purity. The feed is liquid entering the filter module. The retentate is material remaining on the feed side, often returned to the reservoir in a recirculating system. The permeate is liquid that crosses the membrane and leaves through a separate path. Follow these paths on the actual module’s port diagram before interpreting the collected fractions.

    A membrane’s molecular weight cutoff, or MWCO, is based on retention of test molecules. It is not a direct EV-size limit or a guarantee that all EVs remain. Membrane properties and sample behavior affect retention. Some soluble material may remain in the retained fraction. Desired material may also be lost to surfaces or the permeate.

    Fouling is buildup that changes membrane flow or retention. Transmembrane pressure is the pressure difference that drives flow across the membrane. Crossflow, pressure, temperature, sample composition, and processing time affect performance. A pump speed alone is insufficient to describe the operation. Record the measurements supported by the system and use the validated limits for that module and sample.

    Concentration reduces the sample volume by collecting permeate. The volume-based concentration factor is starting volume divided by final retentate volume. Recovery is the amount of a selected material obtained at the end divided by its starting amount. These calculations answer different questions.

    tangable flow.pngWorked example: volume reduction and particle recovery

    A clarified preparation starts at 250 mL with a measured particle concentration of 2.0 × 10⁸ particles/mL. Its final retentate is 25 mL with 1.6 × 10⁹ particles/mL. Assume that both measurements use a comparable validated method and represent the mixed samples.

    Volume concentration factor = 250 mL ÷ 25 mL = 10.

    Starting measured particles = 250 × 2.0 × 10⁸ = 5.0 × 10¹⁰ particles.

    Final measured particles = 25 × 1.6 × 10⁹ = 4.0 × 10¹⁰ particles.

    Measured particle recovery = (4.0 × 10¹⁰ ÷ 5.0 × 10¹⁰) × 100 = 80%.

    The liquid volume decreased tenfold, while measured particle concentration increased eightfold. The particle result does not establish EV-specific recovery unless the measurement distinguishes EVs from other particles. A volume change alone would not have revealed the 20% difference in measured total particles.

    Diafiltration exchanges solution around retained material by adding fresh buffer while permeate is removed. In constant-volume diafiltration, the retentate volume remains approximately unchanged. One diafiltration volume, or DV, equals that retentate volume. For a 25 mL retentate, adding 75 mL of buffer while removing approximately 75 mL permeate supplies three DVs. Freely permeating small components become less concentrated, but complete removal is not guaranteed.

    A rinse after the sample has been removed is a recovery or cleaning step, not the same process as exchanging buffer around a recirculating sample. Label recovery rinses separately so that a later assay can be linked to the correct fraction. Keep starting material, final retentate, and selected permeate portions when the analysis plan requires a mass balance or investigation of loss.

    Attribution: "Tangable Flow" AI-generated using NotebookLM. Modified by Emalee Mackenzie, Irvine Valley College. Licensed under CC BY-NC-SA


    This page titled 14.4: Tangential Flow Filtration was last modified on Wed, 21 Oct 2026 01:35:36 GMT and is shared under a CC BY-NC-SA 4.0 license and was authored, remixed, and/or curated by Emalee MacKenzie.