After completing this chapter, you should be able to:
- Describe the roles of the sample, method, reference materials, calibration standards, and controls in an assay.
- Explain how sampling and sample preparation can affect a result.
- Distinguish accuracy, precision, bias, and random variation, and calculate basic measures of performance.
- Explain selected analytical method validation characteristics.
- Organize data in labeled tables and suitable graphs.
- Use an appropriate calibration equation to estimate an unknown within the validated range and account for dilution.
An assay is a procedure used to evaluate a defined characteristic of a sample. Examples include identity, concentration, biological activity, or a particular impurity. Specificity is usually a performance characteristic of a method rather than a property being measured in a product.
A quantitative assay reports a numerical value, such as concentration. A qualitative assay reports a category, such as detected or not detected under defined conditions. Some assays are semiquantitative and provide an estimate or range rather than a precise numerical result.
For example, a cholesterol test can report a concentration. Many routine urine pregnancy tests detect human chorionic gonadotropin (hCG) and report a qualitative result according to a threshold. A negative result does not prove that no hCG is present; the level may be below the test’s detection capability or the sample may have been collected too early.
Assays support research, development, manufacturing, and quality control. They may measure an active ingredient, detect contamination, evaluate degradation, or help determine whether a material meets specifications. A laboratory purity or potency test does not by itself establish clinical efficacy in patients.
- 15.1: Categories of Quality Tests
- This page covers the assessment of materials through key characteristics like appearance and purity. It details identity tests (chromatography, spectroscopy), purity and contaminant detection tests (endotoxins, mycoplasma), and methods for measuring analyte quantities and evaluating biological activity (potency tests). Stability studies are important for monitoring product quality over time, with testing requirements differing based on dosage forms and administration routes.
- 15.2: Selecting and Establishing a Method
- This page outlines methods for analysis from various reputable sources and emphasizes the selection of appropriate techniques based on sample characteristics. It stresses the importance of validating procedures under actual conditions and following Standard Operating Procedures (SOPs) to maintain integrity and compliance with regulations, while cautioning against unplanned repeat tests for favored results.
- 15.3: The Sample
- This page highlights the significance of representative sampling in examinations, focusing on effective selection methods to minimize bias and achieve a sample that reflects the wider population. It underscores the importance of random selection, and discusses various methods, including stratified sampling, particularly in clinical contexts. The text stresses careful planning and documentation of procedures and factors that can influence results, especially concerning solids and liquids.
- 15.4: Sample Preparation and Recovery
- This page highlights the critical role of sample preparation in analytical testing, covering processes such as thawing, mixing, and filtration. It emphasizes the need for consistent procedures, documentation, and understanding variations. Key concepts include spiking samples to assess recovery rates, calculating recovery percentages, and ensuring matrix matching with established criteria for method performance indicators.
- 15.5: Reference Materials Standards and Controls
- This page covers the properties and classifications of reference materials, highlighting certified reference materials and the significance of using Standard Reference Materials (SRMs) from NIST and USP standards. It outlines the various types of calibration standards and controls used in assays, emphasizing their unique purposes.
- 15.6: Analytical Method Validation
- This page discusses analytical method validation, highlighting its importance in ensuring procedures meet intended purposes by assessing performance characteristics appropriate to their application. It outlines the FDA's ICH Q2(R2) guidance for creating a validation plan that includes intended use, sample matrix, performance studies, and acceptance criteria.
- 15.7: Measurement Error and Accuracy
- This page discusses measurement error, highlighting the difference between measured and reference values with concepts of accuracy, bias, and precision. It categorizes errors into gross mistakes, systematic errors, and random errors, emphasizing the need for good technique and calibration. Key practices to improve accuracy are mentioned, along with descriptive calculations like absolute percent error and relative bias, noting the limitations of percent calculations at zero reference values.
- 15.8: Precision
- This page covers precision in measurements, detailing repeatability, intermediate precision, and reproducibility in labs. It warns that a tight clustering of results can still be biased, and a close mean does not guarantee reliability. Key statistical terms including sample standard deviation, population standard deviation, and coefficient of variation are defined with corresponding formulas.
- 15.9: Detection and Quantitation Limits
- This page covers essential concepts in analytical chemistry, focusing on the limit of detection (LOD) and limit of quantitation (LOQ), noting that LOD is always lower than LOQ. It discusses the importance of sensitivity in analytical calibration and clinical diagnostics, and outlines methods for evaluating LOD and LOQ using statistical techniques, highlighting the need for experimental confirmation of these estimates.
- 15.10: Specificity and Selectivity
- This page emphasizes the importance of specificity and selectivity in analyte measurement, considering potential interferences from sample matrices and impurities. It notes that methods may be selective but not entirely exclusive, requiring confirmation through additional techniques for unexpected outcomes. An example is provided with total protein assays that can detect diverse proteins but do not specifically identify therapeutic ones.
- 15.11: Calibration Model Linearity and Range
- This page emphasizes the necessity of using suitable calibration models in assays, particularly cautioning against linear models in immunoassays where sigmoidal models are often more appropriate. It highlights the significance of having a reportable range for reliable results and validating dilution steps. The page warns against relying on linear equations in saturation regions, advocating for validated nonlinear models or proper dilution into a confirmed range for accurate assay outcomes.
- 15.12: Robustness and Intermediate Precision
- This page explores robustness in procedure performance assessment against slight, intentional parameter changes. It emphasizes the necessity of empirical studies to establish operational controls. Although pH variations affect retention time in chromatography, they do not inherently signify robustness issues.
- 15.13: System Suitability
- This page discusses system suitability tests, which validate the measurement systems’ adherence to necessary criteria for accuracy. It covers aspects such as precision, resolution, peak shape, and response, and emphasizes the importance of following approved protocols for timing, calibration, and equipment.
- 15.14: Organizing Quantitative Data
- This page details the key elements of presenting data in tables, such as clear titles and units, and covers calibration graphs with standard concentration as the independent variable. It emphasizes the importance of retaining original data for educational purposes, noting that the provided CRP data is not clinically validated. Furthermore, it stresses the accurate reporting of all observations and their variability.
- 15.15: Graphs
- This page covers data visualization techniques, highlighting scatter plots for paired data and histograms for distributions. It stresses the importance of labeling axes, using appropriate scales, and ensuring clarity. Advice is given on creating effective titles, displaying measured points, and using error bars. It notes that graphs may not represent the full coordinate plane and discusses considerations for connecting points in plots.
- 15.16: Standard Curves and Linear Regression
- This page emphasizes the significance of calibration curves in analytical methods for linking concentrations to measured responses. It covers the necessity of using suitable standards, treatments, and sample preparations, while managing matrix and timing effects. Key concepts include interpolation and extrapolation via a linear model, along with the importance of correlation coefficients (r) and R² values for model assessment.
- 15.17: Unknown and Dilution Example
- This page explains calculating solution concentration from absorbance using a fitted line, illustrating that an absorbance of 0.500 corresponds to a concentration of 561 µg/mL after adjustment for dilution. It highlights the need for correct dilution factors, the limit for reporting concentrations (e.g., not exceeding 1.728 absorbance), and the importance of re-testing diluted samples.
- 15.18: Chapter Summary
- This page discusses the importance of quality assays for evaluating sample characteristics, highlighting quantitative and qualitative methods. It emphasizes the need for reliable sampling, calibration standards, and controls for performance verification. Analytical method validation is addressed through assessments of accuracy, precision, and specificity.