Preparation may involve thawing, mixing, extraction, dilution, filtration, centrifugation, or removal
of interfering substances. Each step can affect the target analyte and the sample matrix, the other materials surrounding it. Adsorption, degradation, incomplete extraction, or dilution errors can change the result.
A preparation effect is systematic when it creates a consistent directional bias. Variable loss can also increase random variation. Use preparation conditions shown to be suitable, document timing and storage, and assess stability during handling.
Prepare samples, standards, and controls consistently where the method requires comparable treatment. They need not always have identical composition or every identical preparation step. Appropriate matrix matching, independent controls, and correction approaches depend on the method and validation evidence.
A spiked sample contains an added known quantity of analyte. A recovery study compares the measured increase with the amount added, accounting for the analyte already present. Adding a spike before extraction can assess preparation losses, while adding it afterward may help assess matrix-related measurement effects. A spike does not always behave exactly like analyte naturally present in the sample.
Recovery (%) = [(result for spiked sample − result for unspiked sample) / amount added] × 100
All quantities must be expressed on a comparable basis, with relevant dilution and volume changes included. Define acceptable recovery before testing and interpret it together with other method performance evidence.