A reference material has properties sufficiently well established for a particular measurement purpose.
A certified reference material includes specified documentation, assigned property values, and associated uncertainty and traceability information as applicable. Not every reference material is a highly purified chemical.
NIST provides Standard Reference Materials (SRMs) for many measurement applications. USP also supplies reference standards for specified compendial uses. Check the intended use, certificate or instructions, storage, expiration or valid-use conditions, and handling requirements. Do not assume every reference standard from every source has the same certification status.
A calibration standard has an assigned value used to establish the relationship between response and quantity. A control checks whether the test system performs acceptably. One material can sometimes serve more than one purpose, but independent control preparation can reveal errors that would be missed if the same mistake affected both calibrators and controls.
Choose controls appropriate to the assay. Qualitative tests often include positive and negative controls. Quantitative assays may use control samples at low, middle, or high levels, blanks, system-suitability tests, or other checks. There is no single control design that applies unchanged to every assay.
A positive control should give the expected response when the target or activity is present. It may be qualitative or have a known concentration. A negative control should give the expected result in the absence of the target or relevant activity. Depending on the method, this might be a matrix blank, a reagent blank, or a no-template control.
An extraction control assesses preparation steps; a reagent blank checks background introduced by reagents; and an internal control can help detect inhibition or other failures within individual samples. These controls answer different questions and are not necessarily interchangeable.
Define control acceptance criteria and what to do when they fail. A failed control may invalidate a run or affect particular results, depending on the method and cause. Investigate before interpreting affected samples. A passing control does not prove that sampling, identification, and all other steps were correct.
This page titled 15.5: Reference Materials Standards and Controls was last modified on Mon, 09 Nov 2026 00:11:25 GMT and is shared under a CC BY-SA 4.0 license and was authored, remixed, and/or curated by Emalee MacKenzie.