    5.3: Experimental and Instructions

    Materials

    • Equipment:
      • Nicolette iS50 FT-IR Spectrometer
      • Long Path Length Gas Cell with windows that are transparent to IR radiation (KBr windows)
    • Reagents:
      • Hydrogen Chloride 99+%: Aldrich Chemical Company; Cat # 29,542-6; Cas # [7647-01-0]; FW = 36.46; safety information can be obtained from: Sigma Aldrich (Click).
    HAZARD: This experiment uses concentrated HCl (strong acid)! 

    The concentrated acid used in this experiment is a hazard. Go to the linked Safety Data Sheet and read the safety information related to these compounds:

    • Hydrogen chloride is colorless, pungent, and corrosive .
    • Contact will cause burns to the skin, severe burns to the eyes, and burns to the respiratory system if inhaled. 
    • Use gloves, protective eyewear, and avoid inhalation throughout this experiment.
    USE CARE: Gas Cells are FRAGILE and the KBr Windows are FRAGILE andare destroyed by water!

    The cell is constructed of a glass body with KBr windows at each end. Ensure that the salt windows are kept free of moisture at all times. Please do not touch these windows.When not in use the IR gas cell must be stored in the desiccator.

    The Group Notebook (ELN) Template is available through Google Classroom. This template is provided as a guide so that you know what data and information should be kept in a notebook. You may keep your personal notebook in any way you'd like, but you should translate your notes into this template for the group submission.


    Part I: What can an FTIR Spectrum tell us about a diatomic gas molecule?

    A: Computational prediction of vibrational frequency and bond length

    1. Use a computational chemistry program to determine the computationally-predicted vibrational frequency and bond length of hydrochloric acid gas.
      1. Perform Equilibrium Geometry Calculation: Specifically, use density functional theory (DFT) to calculate the equilibrium geometry of HCl gas, determine its vibrational frequency (or frequencies), and the equilibrium bond length (internuclear distance). 
      2. Record Computational Results: Record the predicted IR spectrum in your group notebook.  Also record the predicted frequency of the vibration, the predicted bond length, and any other information you feel is worth recording.

    Part IB: Preparation of sample and collection of FTIR spectrum

    SEE HAZARD and CAUTION STATEMENTS above.

    Use the following method (see video) of preparing a gas cell.  Start with the cell filled only with the gases from atmosphere: collect a spectrum of the background. Then, add one drop of HCl, allow it to evaporate in the sealed cell, and use the cell to collect a spectrum of HCl gas.

    Prepare the cell with background gas  
    1. Put on clean, dry gloves before handling any components of the gas cell.
      Avoid touching the Potassium Bromide (KBr) windows with bare hands, and never expose them to water, as they are made of salt and will degrade upon contact with moisture.
    2. Gather all required materials, including:
      • One Pyrex gas cell body
      • Two Potassium Bromide (KBr) windows
      • Two blue end caps
      • Four O-rings
      • Two septa
      • pH paper
      • Concentrated hydrochloric acid
      • Glass pipettes
      • Wipes
    3. Insert a small piece of pH paper into one of the inlets of the gas cell.
      Fold the pH paper as needed so that it fits easily inside the inlet.
    4. Cap the inlet containing the pH paper with a septum.
      The septum does not need to fold down or form a tight seal; it simply needs to securely plug the opening.
    5. Prepare the end caps by placing one O-ring into the groove inside each end cap.
      Use your thumb to press the O-ring evenly into the groove.
    6. Place one Potassium Bromide window into each end cap.
      If a window has one side that appears etched or foggy, orient that side so it faces inward toward the cell body.
      Keep the windows dry at all times.
    7. Place a second O-ring on top of each Potassium Bromide window.
    8. Gently screw the body onto the end caps without distrubing the windows and o-rings.
      Do not overtighten. The Potassium Bromide windows are fragile and can crack or crush if excessive force is applied.
      Tighten only until slight resistance is felt.
    9. Collect a background spectrum on the sealed gas cell assembly using the FTIR instrument.
    Introduce HCl into the cell
    1. Carefully introduce hydrochloric acid into the cell.
      Using a glass pipette, add one drop of concentrated hydrochloric acid through the inlet.
      Ensure that the liquid does not contact the Potassium Bromide windows.
      If any acid is spilled, clean it immediately and change your gloves before proceeding.
    2. Wipe the outside of the inlet to remove any residual acid, then cap the inlet with a septum.
    3. Gently move and rotate the cell so that the liquid drop moves away from the Potassium Bromide windows and toward the center of the cell.
      Rotate the cell slowly to spread the liquid around the inner circumference, increasing surface area and promoting evaporation.
    4. Monitor the pH paper.
      The cell is ready when the pH paper indicates the presence of acid (e.g., a color change).
    5. Collect the FTIR spectrum of gaseous hydrochloric acid.
    Disassemble the cell and clean up
    1. Disassemble the cell carefully after data collection.
      Remove at least one septum before unscrewing the end caps.
      Unscrew the end caps slowly, ensuring that the acid does not contact the Potassium Bromide windows.
    2. Set the end caps aside carefully and keep the windows dry.
    3. Remove the pH paper using a clean glass pipette and tap it out of the cell body.
    4. Change your gloves before handling the Potassium Bromide windows! Transfer the windows to their designated storage location inside a desiccator.
    5. Flush all remaining cell components thoroughly with water and dry them with a towel.
    6. Wipe bench spaces that may have acid residue with a wet towel, followed by a dry towel.
    7. Dispose of all acid-contaminated waste in the appropriate waste container.

      


    Data Analysis for HCl FTIR spectrum for Part I 

    After collecting your first spectrum, please email your results to yourself, your group, and your instructors. Then assemble your group and begin group discussion and analysis of the data. The files linked at the top of this page will guide you to analyze the spectrum, derive mathematical models to fit the data, and calculate molecular parameters from the fits. (links copied here for convenience: Shared Files for Duke Students). Your instructors will be on hand to help guide you if you get stuck.


    Part II: How does instrument resolution effect the certainty of derived parameters?

    In Part II, you will collect the spectrum of HCl gas at different instrument resolutions to determine how instrument resolution affects what the IR spectrum can show us, and how instrument resolution affects the uncertainty in the results.

    If you finish with part I early, you may be able to collect Part II data in the first lab period. Otherwise, data collection for Part II will happen in the second week of this module.

     


    This page titled 5.3: Experimental and Instructions was last modified on Sat, 19 Sep 2026 15:27:09 GMT and is shared under a CC BY-NC-SA 4.0 license and was authored, remixed, and/or curated by Kathryn Haas.