Pre-Lab Assignment
📅 Due: This pre-lab assignment is due prior to the Week 1 Dry Lab meeting.
🎯 Purpose: Prepare for the Week 1 Dry Lab meeting and Wet Lab session by selecting your beverage, planning your calibration strategy, and collecting background information for key parts of your scientific manuscript.
- Propose at least four beverage samples to examine, and research their caffeine content:
(This is just an initial proposal, you will refine it with your team during the Dry Lab meeting.)- Select at least four beverage samples that you would be interested in examining.
- Research their approximate caffeine concentration using reputable sources (e.g., USDA FoodData Central, peer-reviewed publications, manufacturer data).
- Save the source(s) to your Zotero library and cite in ACS style.
- Propose a method to quantitate caffeine content of your beverage sample.
- Be sure you propose methods that are accessible to your team.
- Justify your proposal with evidence from at least one peer-reviewed literature source.
- Cite sources in ACS Style.
- Draft Introduction (partial)
- Write 1–2 paragraphs of background for your Introduction section. (This is just a first draft.)
Include:- Brief explanation of caffeine’s significance (chemical properties, psychoactive nature, health considerations), with citations.
- Why accurate caffeine quantitation matters (public health, labeling regulations, quality control), with citations.
- The structure of caffeine (create a figure using ChemDraw) and its chemical or physical properties that allow its detection by the proposed method.
- Include at least three references in ACS style.
- Write 1–2 paragraphs of background for your Introduction section. (This is just a first draft.)
- Proposed plan to create calibration standards & to create diluted beverage samples
(This is just a first draft, you will refine it with your team during the Dry Lab meeting.)- Select the tools (glassware, pipettes) and create a dilution plan for creating caffeine calibration standards in the approximate range 0.005 – 0.04 g/L (the range does not need to be exact).
- Determine the dilution factor needed to bring your beverage sample into the middle of your calibration range. Keep in mind that your highest calibration point should be between 1.5 to 3 times more concentrated than that of your sample.
- Present your plan in a clear, stepwise outline or table.
- Prepare a draft data table template
(This is just a first draft, you will refine it with your team during the Dry Lab meeting.)- Create a table to record your UV-Vis absorbance readings for calibration standards and beverage samples.
- Include columns for:
- Sample ID
- Caffeine concentration (g/L)
- Absorbance at λmax
- Replicate #
- Notes/Observations
Submit your pre-lab assignment on Canvas, and be prepared to justify your plan for solution preparation to your team.
👥 Dry Lab Group Discussion
The purpose of today’s Dry Lab is to share and critique your individual pre-lab plans for calibration and sample preparation. By the end of this session, you should have considered multiple approaches, anticipated challenges, and refined your own plan for Wet Lab. You will also begin drafting sentences for the Methods section of your manuscript.
Instructions:
- Begin the session by grouping up and assigning roles (scribe, reporter, spokesperson, manager)
- The Manager will start a timer and encourage group discussion and consensus.
- The Scribe will record the group work.
- The Spokesperson will summarize your group’s discussion and questions to the instructors.
- The reporter will submit the group record at the end of class.
Goals for Dry Lab:
- Work as a team to refine your calibration plan and sample dilution plan so you can start preparing solutions right away in Wet Lab.
- Work with your team to make a plan for dealing with technical issues during data collection.
- Begin writing the Methods section of your manuscript.
- Calibration Choices: What choices did you each make when designing the plan for your calibration standards?
- What choices did you each make when designing calibration standards (concentration range, number of points, dilution method)?
- What tools did you select, and why?
- How much of each standard will you each produce? Will your volumes provide enough material for UV–Vis (~3 mL), HPLC (~2 mL), and filtering?
- Share your plans for creating the calibration standards. Is the strategy reasonable considering the necessary volumes and the tools available?
- After discussion, what adjustments (if any) will you make?
- Sample: Each group member should choose a unique sample. No two group members should use the same exact beverage.
- Why do we dilute beverage samples instead of analyzing them directly?
- Compare each member’s dilution strategy for the chosen beverage:
- Is it reasonable?
- Is the anticipated diluted sample concentration within the calibration range?
- What challenges might arise with your particular beverage (e.g., temperature, carbonation, particulates, solubility)?
- Validation & Troubleshooting
- What will you do if your standards give absorbances outside the quantitative range for UV-vis (0.2–0.8 au)?
- Are shorter path-length cuvettes realistic to use, or is further dilution more practical? Do you need to collect UV-vis data on every standard to confirm accurate sample preparation?
- Do you have enough solution to accommodate repeats or adjustments?
- Data Collection and Sharing
- Please see the canvas page describing the CHEM401L Data Drive (CHEM401L Data on Google Drive).
- How will your group share data files (UV–Vis scans, calibration results) to make sure everyone has multiple examples for their reports?
- What naming conventions or folders will you use to avoid losing or mixing up files?
- Writing: While this discussion is fresh in your mind, write a few sentences that can support the Methods section of your manuscript.
- Draft one sentence describing your calibration preparation.
- Draft one sentence describing your beverage preparation.
- Draft a one-sentence description of how you will validate your calibration.
🧪 Wet Lab Agenda
Goals: (1) Prepare a beverage sample and analyze it with your chosen method(s). (2) Prepare a set of caffeine calibration standards using your preferred method and tools, and validate your calibration solutions.
*Do not throw away any solutions until the very end of this 3-week module! Plan to store your solutions in sealed vials in the refrigerator for the entire 3 weeks.
- Turn on the instruments and allow an appropriate worm up period.
- If using a Cary UV-vis: Validate the instrument after warm-up, and before using it.
- If using a Spec 20, be sure to "blank" often to avoid issues with lamp intensity drift.
- If using HPLC, begin column equilibration and run two blanks after lamp warm up and before injecting sample.
- If using GSCM, run two blanks before injecting sample.
- If using another instrument, consult your instructor.
Materials required:
- Caffeine (99%. powder)
- Ultrapure water
- Analytical balance (no more than 0.5 mg uncertainty)
- Caffeine Validation Control solution (provided by TAs)
- Equipment for preparing solutions
You have access to:- volumetric flasks to contain of 5, 10, 25, 50, 100, 250, 500, and 1000 mL
- volumetric pipete to deliver volumes of 1, 2, 3, 4, 5, 10, 15, 20, 25, 50, and 100 mL
- automatic pipetters to deliver between 100-5000 mL
A. Example protocol to prepare a stock solution of 0.1 g/L caffeine
- Quantitatively weigh out between 8.0 to 10.0 mg of caffeine. The caffeine can be found on the shelf near the weigh station area.
- Quantitatively transfer the caffeine into a clean 100 mL volumetric flask. Rinse the weigh boat with ultrapure water to be sure all powder is rinsed into the flask.
- Dilute to the volumetric mark with ultrapure water. The stock solution will have a final concentration of ~ 0.1 g/L. Calculate the true concentration based on the actual mass of caffeine used.
(1) You must use quantitative technique to transfer the caffeine from the weigh paper to the receiving flask. After transferring the powder to the flask, rinse the weigh paper carefully with water, and capture the rinse with the flask to be sure that no trace powder is left behind.
(2) There is a very small amount of caffeine powder, but it takes some time to dissolve completely. Be sure that caffeine is completely dissolved in about 50 mL of water before diluting to the volumetric mark and before proceeding to the next step. There is a sonicator in the lab that you can use to help break up small caffeine particles and aid in dissolution.
B. Prepare calibration standards
Write out a specific plan for doing this before you get to lab (see pre-lab assignment). Carry out a serial dilution to obtain a series of calibration standard solutions that are are approximately evenly spaced (in terms of their concentrations). The highest concentration should be about \(0.04 \pm 0.01\) g/L, and the lowest concentration solution should be approximately \(0.005 \pm 0.001\). Make enough of each solution so that you have at least 10 mL of each solution to play with. You should use ultrapure deionized water to make all solutions and dilutions. Ensure adequate mixing in each diluted solution before using it to create the next.
Each student is responsible for bring about 50 mL of at least one unique caffeinated beverage sample for testing. All together, the team is responsible for bringing four different (unique) beverage samples. Beverages used in lab should not be consumed.
Beverages and beverage samples must be stored in the refrigerator to avoid microbial growth. Warm all liquids to room temperature before measuring volumes!
Dilute the beverage samples to reach a concentration that can be quantitated by your calibration range (based on the expected caffeine content of the beverage, and within the range of the calibration standards).
Example beverage preparations:
- A Coffee, which is estimated at ~ 0.4 g/L caffeine.
- Pipette 5 mL of coffee into a clean and dry 100 mL volumetric flask and dilute to the mark with clean deionized water.
- Filter the sample using the provided filter.
- First, rinse the filter three times with 3 mL clean deionized water.
- Rinse the filter with the sample solution by filtering the first 1-2mL of the sample into the waste beaker.
- After rinsing the filter apparatus with sample solution, filter the sample directly into a clean vial and label the vial.
- A Tea, which is estimated at ~ 0.2 g/L caffeine.
- Pipette 10 mL of tea into a clean and Dry 100 mL volumetric flask and dilute to the mark with clean deionized water.
- Filter the sample using the provided filter.
- First, rinse the filter three times with 3 mL clean deionized water.
- Rinse the filter with the sample solution by filtering the first 1-2mL of the sample into the waste beaker.
- After rinsing the filter apparatus with sample solution, filter the sample directly into a clean vial and label the vial.
- A Soft Drink Sample, which is estimated at ~ 0.1 g/L caffeine.
- If the soft drink you selected is carbonated, decarbonate the soft drink by pouring it back and forth between two beakers until the bubbles cease.
- Pipette 10 mL of the soft drink into a clean and Dry 50 mL volumetric flask and dilute to the mark with clean deionized water.
- Filter the sample using the provided filter.
- First, rinse the filter three times with 3 mL clean deionized water.
- Rinse the filter with the sample solution by filtering the first 1-2mL of the sample into the waste beaker.
- After rinsing the filter apparatus with sample solution, filter the sample directly into a clean vial and label the vial.
Analysis of caffeine sample
Use the method(s) of choice to analyze your sample and the control. After collecting this data, compare the sample to the control, and assess the utility of the method(s) for performing quantitative analysis of caffeine. Some things to consider:
- Are the features attributed to caffeine specific to caffeine, or are there matrix interferences that will prevent conclusive quantitation?
- If there are technical issues, what strategies can you use to overcome the problem?
- How confident are you that this method will be useful for quantitation of caffeine in this specific group of samples?
- Are other methods available that might give more reliable quantitation of caffeine in these samples? (Consider different methods on the same instrument, and methods using a different type of instrument.)
Validation of method and calibration standards
If you are confident that the method chosen is promising, then prepare to validate your method.
If possible, use the method to measure the response to the blank, the calibration standards, the validation control, and the samples. If possible during this lab period, construct your calibration curve, evaluate the quality of your data and the fit, and when you feel that you have a useful data set, use the data to determine if the method is valid. Otherwise, plan to make progress on this next week.
Keep in mind that one reason that a method fails validation is due to improperly prepared calibration standards. Prepare for this possibility, and the fact that you may need to prepare your solutions again if they are not appropriately prepared the first time.
📘 Notebook
Record detailed notes about your lab work in your notebook each time you are in lab. Also include the data analysis and responses to the reflection prompts*. Your notebook should be submitted on canvas prior to leaving Wet Lab each week - whatever state it is in at the end of lab is the state in which you should submit your work. If you are unable to complete data analysis and the reflection prompt within the period, it should be completed before the next lab period, and submitted with your next notebook submission.
- What have you learned from your experiments about the method(s), and will it (they) be sufficient for quantitation of caffeine in these samples?
Justify your response with evidence. - Dealing with unexpected challenges: During this experiment, what unexpected challenges or technical issues did you experience with measuring the absorbances of your calibration solutions? What strategy did you use to address this issue, and was it effective?
- Considering the data you collected, what is the "best" wavelength for measuring caffeine concentration? Justify your response based on at least two of the following:
- Beer's Law
- Sensitivity of the signal to changes in concentration
- Signal to noise
- Relative error in the absorption measurement