An experimental image cannot reveal all sources of background on its own. Use controls selected for the labeling scheme and question. Process comparison samples consistently and image them using appropriate matching settings. Controls help identify specific problems; no single control answers every question.
|
Control |
Question it helps answer |
Important limit |
|
Unstained prepared sample |
Is there signal without added label? |
Does not test antibody specificity |
|
Secondary antibody only |
Is signal present without the primary? |
Does not validate primary binding |
|
Known positive sample |
Can the method detect the target? |
Does not establish experimental identity |
|
Negative biological sample |
Is signal present without expected target? |
Must be appropriate for the target |
|
Single-label samples |
Does a label appear in another channel? |
Does not assess all background sources |
An unstained control should undergo the relevant preparation steps when the purpose is to evaluate preparation-related background. A secondary-only control applies to indirect antibody labeling. It can reveal signal that appears without the primary antibody, but it cannot demonstrate that the primary antibody recognizes only the intended target. [5]
Autofluorescence comes from the specimen without an added label. Bleed-through occurs when one label’s signal appears in another label’s channel. Single-label controls help reveal that overlap. Colors in a merged image are assigned for display; they do not provide separate evidence that the labels are specific.
A positive control helps determine whether a method can detect the target under the selected conditions. A negative biological control helps assess signal when target expression is absent or reduced. Where appropriate, target knockdown can provide evidence, but incomplete reduction or altered cell health can complicate interpretation. Explain the control’s role and limits.
If the controls fail, an attractive experimental image does not rescue the result. A bright secondary-only control, for example, calls for investigation of background before interpreting experimental brightness as target expression. Preserve failed control images and record what was changed during troubleshooting.