Place the correct coverslip on a clean chamber as instructed. Capillary action draws liquid into the narrow space beneath it. Load the sample at the chamber entrance. Bubbles, incomplete filling, or overflow can change the counting volume or distribution. Reload a faulty chamber before calculating a result.
Use enough magnification and suitable illumination to distinguish cells, dye staining, and grid lines. Scan each chosen square in an organized pattern so that cells are not missed or counted twice. Record clear and blue counts for each square. A total without the individual square counts hides uneven distribution.
A boundary rule specifies how to count cells touching grid lines. One consistent rule is to include cells touching the top or left boundary and exclude cells touching the bottom or right boundary. It prevents shared edges from being counted twice. Apply the same convention to all squares and both cell categories. For a cell touching an excluded edge at a corner, exclude it. If cells overlap heavily, make an appropriate additional sample dilution and recount. If very few cells are present, count more defined areas or use an approved method to improve the measurement. Record all changes. A calculation cannot correct uncertainty caused by guessing how many cells are hidden in a crowded region.