Opening the book…
A hemocytometer is a counting chamber with an engraved grid and a defined chamber depth. The correct coverslip establishes that depth. In a standard Improved Neubauer chamber, one large square has an area of 1 mm² and the chamber is 0.1 mm deep. The liquid above one large square therefore occupies 0.1 mm³, which equals 0.0001 mL.
Counting cells in this small volume estimates how many would be present in 1 mL. Because 1 mL contains 10,000 volumes of 0.0001 mL, multiply the average count per large square by 10,000. This is a volume conversion, not a factor chosen to make the answer larger. A sample dilution must be corrected separately.
Figure 8.1 A simplified standard counting grid. The four shaded corner squares are each 1 mm by 1 mm. The center square has a different subdivision pattern. Always identify the area actually counted.
The chamber contains squares of different sizes, so each type represents a different volume. In this chapter’s four-corner-square method, count four large squares and divide the combined count by four. Multiply that average by 10,000. Applying 10,000 directly to the combined four-square count would ignore the larger volume counted.
A more general relationship is concentration = cells counted ÷ volume counted. This works for any known counting volume. Use the device dimensions and counting plan provided for the actual chamber. Disposable slides and automated instruments may use different volumes or perform the conversion internally.
"Counting cells.m4v" LafayetteChBE https://www.youtube.com/watch?v=8nmh3yYuJmM Standard YouTube License. No changes made.
This page titled 10.3: How the Hemocytometer Works was last modified on Wed, 21 Oct 2026 01:35:29 GMT and is shared under a CC BY-NC-SA 4.0 license and was authored, remixed, and/or curated by Emalee MacKenzie.