    13.11: Analytical Laboratory Documentation

    Biotechnology method, sample, and test documentation.pngAnalytical records describe the sample, test method, work performed, and result. An environmental laboratory may measure water pH or turbidity; a clinical laboratory may measure blood components; a pharmaceutical QC laboratory may assess drug potency or impurities. Applicable requirements differ, but results need adequate supporting evidence.

    Method information can include the method identifier and version, test purpose, relevant measurement range and limits, source, suitability, and verification or validation information. A detection limit is different from an acceptance limit; specify which limit is being reported.

    Sample information includes a unique identifier, source, collection method and collector, collection and receipt times where relevant, storage, condition, and disposition. Link the tested portion to the original sample.

    Performance information includes the test date, analyst, instrument, reagents and lots, method, raw data, calculations, controls, results, and review. For finished pharmaceutical testing, 21 CFR 211.194 describes laboratory record requirements.


    This page titled 13.11: Analytical Laboratory Documentation was last modified on Sun, 08 Nov 2026 21:08:19 GMT and is shared under a CC BY-SA 4.0 license and was authored, remixed, and/or curated by .