15.17: Unknown and Dilution Example

Quality and Regulatory Compliance in Bioscience · LibreTexts

0%

Contents

    Text

    Size
    100%
    Theme
    Typeface
    Line length
    Spacing

    Read aloud

    How

    Speed: 1×
    Light up
    Highlight

    Symbols: Mulberry Symbols, © 2018–2026 Steve Lee, CC BY-SA 4.0 (licence, the set), the site's own copies.

    From LibreTexts, under CC BY-SA 4.0, as the source states it; its equations are written in TeX, which the PDF shows typeset. This page at LibreTexts.

    15.17: Unknown and Dilution Example

    Suppose a teaching sample gives an absorbance of 0.500 and all required method criteria are assumed to pass. Concentration calculation teaching infographic.pngUsing the fitted line:

    x = (0.500 − 0.01133083) / 0.00435188 = 112.29 µg/mL

    This estimates the concentration of the solution actually measured. Report a rounded result appropriate to the method’s precision; the extra digits here show the calculation, not proven analytical certainty.

    Dilution factor and concentration guide.pngIf the measured solution was prepared by diluting 1.00 mL of sample to a final volume of 5.00 mL, the dilution factor is 5.00. Under the assumed valid dilution procedure:

    Original concentration = measured concentration × dilution factor

    Original concentration = 112.29 × 5.00 ≈ 561 µg/mL

    A dilution factor is final volume divided by sample volume, not diluent volume divided by sample volume. Keep the units and any additional preparation factors consistent.Out-of-range assay response guidance.png

    An absorbance of 1.728 exceeds the highest listed teaching standard response of 1.398. Do not report a concentration by extending this teaching line. Under a suitable method, dilute the sample to bring its response within the validated working interval, retest with required calibrators and controls, and account for the dilution. A serial dilution is one option, not a universal requirement.

    Values below the validated lower limit also need the method’s reporting approach; concentrating a sample or changing preparation cannot be assumed valid. Preserve all results and investigate unexpected or failing controls through the quality system.


    This page titled 15.17: Unknown and Dilution Example was last modified on Mon, 09 Nov 2026 00:49:37 GMT and is shared under a CC BY-SA 4.0 license and was authored, remixed, and/or curated by Emalee MacKenzie.