Equipment logs document relevant use, status, cleaning, calibration, maintenance, repair, and problems. They may be paper or electronic. Identify the equipment uniquely and connect the entry to the relevant activity, person, date, and result. The procedure determines which events must be recorded.
Temperature logs are used for equipment such as refrigerators, freezers, and incubators. Record actual readings with units, defined acceptable limits, and any required minimum or maximum readings. Investigate excursions and document actions and potential effects on stored materials. A log with only the headings Time 1 and Time 2 is insufficient if it does not clearly capture the measured temperatures.
A temperature log example appears in the practice forms. Its blank limits must be set for the equipment and materials; there is no single acceptable temperature range for all storage units.