    Light up
    Highlight

    Symbols: Mulberry Symbols, © 2018–2026 Steve Lee, CC BY-SA 4.0 (licence, the set), the site's own copies.

    From LibreTexts, under CC BY-SA 4.0, as the source states it; its equations are written in TeX, which the PDF shows typeset. This page at LibreTexts.

    15.11: Calibration Model Linearity and Range

    A linear calibration model describes a response that changes approximately in proportion to concentration, allowing for an intercept. A method may instead require an appropriate nonlinear model. Many immunoassays use sigmoidal calibration models, so a standard curve is not necessarily a straight line.

    The reportable range is the interval of results over which the procedure provides suitable performance for its intended use. Relevant dilution or concentration steps can connect a sample’s original level to the instrumental working range. Validate or verify those steps as appropriate.

    A protein assay can lose proportionality at higher levels. Do not use a straight-line equation in a saturated region. Depending on the approved method, dilute into a validated working range or use a suitable, validated nonlinear model. Do not choose a model merely because it gives a favorable value for an unknown.

    Calibration models, working ranges, and saturation.png