    3.6: Ventilator Alarms

    Every ventilator has alarms that are set for the safety of the patient. We have talked a lot about the danger of applying pressure to the fragile alveoli. Using alarms on the ventilator will make sure that the pressures and volumes stay in safe ranges. After setting up your patient and initiating ventilation with your settings, go into your alarm screen and ensure your alarms are appropriate to your settings and patient.

    Red emergency light encased in a metal cage, with green blurred lights in the background.
    Ventilator alarms are a useful tool for keeping you aware of your patient’s status when not present at the bedside. Photo by TBIT, CC0.

    The standard alarms that are set for all ventilation modes are as follows:

    Table 3.6.1 Ventilator Alarms

    Type of Alarm

    Alert

    Significance

    Input Alarms Loss of electric power Activated when the power supply is interrupted while the ventilator is on
    Loss of pneumatic power Activated when either the O2 or air supply pressure is reduced below a specified point
    Control Alarms The set parameters are incompatible Activated when operator makes errors in set up
    Part of the self-test failed Activated during self test prior to initiating ventilation: Ex. Leak in the circuit, or failed flow sensor calibration
    There is something wrong with the ventilator control circuitry Activated when ventilator microprocessor fails to function (usually alarms “vent inop”)
    Output Alarms Volume Alarms High and Low Expired Tidal Volume
    Flow Alarms High and Low Expired Minute Volume
    Time Alarms High and Low Ventilatory Frequency
    Inappropriate Inspiratory Time
    Inappropriate Expiratory Time
    Pressure Alarms High and Low Peak Airway Pressure
    High and Low Mean Airway Pressure
    High and Low Baseline Pressure
    If you would like additional information about the ventilator alarms listed above, check out Ventilator Alarms in Mechanical Ventilation.

    It is important that you ensure your alarms are adequately set for all patients. Do not set your alarms too close to what your normal numbers are. These alarms are your safety net that your patient will not be exposed to volutrauma or barotrauma, but if your alarm ranges are set too tightly, it could cause any patient movement or change to trigger an alarm. Frequent alarms are disturbing for patients as well as the clinician. We also know that if things are alarming too often, it can cause alarm fatigue in healthcare workers and sometimes important alarms can be overlooked. Suggested ranges for alarm settings are below, but always defer to your health centre’s guidance when it differs from these recommendations:

    Table 3.6.2: Alarms and their Suggested Settings
    Alarm Suggested Setting Rationale
    High RR Outline of a coffee cup with steam above it, representing a hot beverage. Patient can wake up or rouse from sedation and increase their RR. This alarm is mostly used for spontaneous modes
    High Pressure A simple black outline of a leaf with visible veins, representing natural elements and organic forms. or
    A simple black silhouette of a cat sitting on its haunches with its tail curled around its body. above your Peak Pressure
    If this limit is hit, it will cut off the breath that is being delivered. This can be very uncomfortable for the patient and cause coughing and asynchrony. Monitor your pressures carefully and try to stay below A simple black and white illustration of a light bulb icon. if possible. This is just the maximum limit.
    Low Pressure Set A simple black silhouette of a cat sitting upright with its tail curled around its body. below PEEP Not set on every ventilator. It is good for sensing a leak or disconnect in the circuit.
    High Volume A simple black silhouette of a cactus with several arms and a rounded base. from your target volume Monitor your volumes breath to breath. This alarm is set wide to avoid alarm fatigue. We as clinicians want to monitor our tidal volumes tighter than this when we are present, but this alarm will come into play when clinicians are not present.
    Low Volume A simple black silhouette of a cat sitting with its tail curled around its body. from target tidal volume Not important in control modes. Usually used for spontaneous modes
    High Minute Volume A simple black and white illustration of a leaf, representing nature or environmental themes. Set widely on purpose. Usually used in spontaneous modes
    Low Minute Volume A simple black silhouette of an apple with a leaf on top. or
    A simple black silhouette of a bicycle. below the MV reading on the ventilator.
    Set widely on purpose. Usually used in spontaneous modes
    Apnea time A simple black silhouette of a tree with a thick trunk and spreading branches. Standard used in adult patients.

    “Ventilator Alarms” from Basic Principles of Mechanical Ventilation by Melody Bishop, © Sault College is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License, except where otherwise noted.