In the 1800s, as the field of cell culture was being developed, researchers began searching for the ideal in vitro environment in which to maintain their cell culture stocks. The earliest cell incubator was a simple bell jar into which a lit candle was placed next to the cell culture. The jar was then placed into a heated oven. Laboratory incubators were first properly introduced during the second half of the twentieth century, when doctors realized that they could be used to identify pathogens from the bodily fluids of patients. During the late 1960s, the first dedicated, commercial CO2 incubator was developed by New Brunswick Scientific (NBS). These incubators maintained a temperature of 37ºC and CO2 concentrations capable of supporting cell growth. The modern incubator is not that different from these first commercial attempts. However, they have had additional improvements added, including better control of CO2, temperature, and humidity levels. Today, the iculture incubator has become an indispensable piece of equipment for the fields of cell biology, microbiology and molecular biology.
Introduction
The major requirement of any cell culture laboratory is the need to maintain an aseptic work area. If possible, a separate culture room is preferred. This room should be small, with no windows that can open, and ideally one door so that air drafts are minimized. The equipment used in culturing cells should be placed inside of this room and their use should be dedicated to cell culture only. Much of the equipment of cell culture is unique and is designed to provide the aseptic conditions critical to successful cell culture.
Cell culture equipment is used to ensure the integrity of the aseptic work area. At the end of this page, you will be able to:
- List the basic equipment of the cell culture lab
- Compare the three biosafety cabinet classes
- Explain how the Class II biosafety cabinet works
- Compare the four types of Class II biosafety cabinets
- Explain how the cell culture incubator works
- Explain the major types of cell counting systems
- Explain how the inverted microscope works
The Basic Equipment of the Culture Lab
The major focus of cell culture equipment is to provide a sterile environment so that contamination of cell cultures does not occur. The typical cell culture lab will contain the following equipment:
- Cell culture hood (i.e., the biosafety cabinet)
- Incubator
- Water bath
- Centrifuge
- Refrigerator (4°C) and freezer (–20°C)
- Cell counting system
- Inverted microscope
- Liquid nitrogen (N2) freezer or cryopreservation container
- Sterilizing equipment (i.e., autoclave)
- Aspiration pump (peristaltic or vacuum)
The Biosafety Cabinet
The cell culture facility requires a place to aseptically manipulate cells. This is achieved using a biosafety cabinet (BSC; i.e. a cell culture hood). The BSC is an enclosed, ventilated workspace for safely working with biological materials that require a defined biosafety level. There are three main types of BSCs, with each providing different levels of protection to the user, the environment, and the biological product. Table \(\PageIndex{1}\) compares the three types of BSCs.
| Class of Biosafety Cabinet | Personnel Protection | Product Protection | Environmental Protection | Use |
|---|---|---|---|---|
| Class I |
Yes Inward air flow through sash opening |
No Unfiltered room air is drawn across the work surface |
Yes Exhaust air is HEPA-filtered |
Not used for bioagents May be used to enclose equipment or procedures with aerosol potential |
| Class II (Type A1, A2, A3, A4) |
Yes Inward air flow through sash opening |
Yes HEPA-filtered air is drawn down onto the work surface and room air is blocked from the work surface |
Yes Exhaust air is HEPA-filtered |
Most common class of BSC Used to handle specimen material, biological toxins, cell and tissue culture, and biohazard materials |
| Class III |
Yes Complete containment of interior work area |
Yes HEPA-filtered air is supplied to the work surface; room air is completely blocked from the work surface |
Yes Exhaust air is double HEPA-filtered |
Provides the highest level of containment for handling of extremely dangerous microorganisms |
It is important not to confuse a BSC with a laminar flow hood. A laminar flow hood draws air in from the environment, passes it through a HEPA filter and then moves the cleaned air across the work surface from back to front, protecting the materials in the hood from contamination. However, this curtain of air is blown out the hood at the worker placed in front of the hood. Therefore, the laminar flow hood provides no protection to the worker using the hood. In contrast, the air flow inside of a BSC is designed to protect the materials inside the cabinet, in addition to the worker seated in front of it.
Class II Biosafety Cabinets
The most commonly used type of BSC in the cell culture laboratory is the Class II cabinet (Figure \(\PageIndex{1}\)). The interior of this class of BSC, including the work surface, is made of stainless steel and can be disassembled for sterilization through autoclaving. The front of the work surface is a large grill that collects air from the environment. A grill located at the back of the work surface collects air from the interior of the cabinet. At the front of the class II BSC is a glass sash that can be raised and lowered. When working in the cabinet, the sash is partially raised to allow for manipulation of the materials inside the hood but limit the entrance of room air with contaminants. A HEPA filter, exhaust manifold and a blower motor are inside the top of the cabinet. Inside the cabinet are visible and ultraviolet (UV) light bulbs. The UV light can be turned on for short periods of time to help sterilize the work surface when it is not in use. A control panel at the front of the cabinet turns on the lights and the blower motor for airflow. It also provides key information about the status of the cabinet.
The Class II BSC creates a unique pattern of airflow within the cabinet that protects both the material inside of the cabinet and the worker using it (Figure \(\PageIndex{2}\)). Contaminated room air is drawn in through the open sash into the front grill. The air flows beneath the work surface and up the back of the cabinet. It enters a HEPA filter at the top of the cabinet where the air is filtered. A portion of the filtered air is exhausted out to the environment and the remainder is blown down over the work surface from the front of the cabinet to the back. This air then flows through the back grill to mix with the room air flowing up to the HEPA filter. The air pattern over the work surface creates a “curtain” to protect the worker and efficiently moves potentially contaminated air out of the work area.
Animation: Airflow Through a Class II Biosafety Cabinet
There four types of Class II biosafety cabinets are classified as A1, A2, B1 and B2, with each type having unique specifications (Table \(\PageIndex{2}\)). Most cell culture facilities use a either a type A1 or A2 biosafety cabinet because they are cheaper to operate since they do not require a hard connection to the exhaust system of the room’s building.
| Type II Classification | HEPA-Filtered Air Recirculation | Ducting | Use |
|---|---|---|---|
| A1 | Most is recirculated over the work surface (70% recirculated; 30% exhausted to room); airflow rate of 75 linear feet per minute | Cabinet is not connected to building ventilation system; filtered air exhausted to environment | Use with biological materials only; do not use with chemicals |
| A2 | Most or all is recirculated over the work surface (70% to 100% recirculated; 0 to 30% exhausted to room); airflow rate of 100 linear feet per minute | Cabinet is not connected to building ventilation system; filtered air exhausted to environment | Use with biological materials; can be used with materials containing minute amounts of volatile chemicals if air is exhausted to environment |
| B1 | Most is exhausted to environment (30% recirculated; 70% exhausted to room through a ducting system); airflow rate of 100 linear feet per minute | Cabinet is directly connected to building ventilation system | Use with biological materials and minute amounts of volatile or toxic chemicals |
| B2 | All is exhausted to environment (0% recirculated; 100% exhausted to room through a ducting system); airflow rate of 100 linear feet per minute | Cabinet is directly connected to building ventilation system | Use with biological materials and small amounts of volatile or toxic chemicals |
The Culture Incubator
Cultured cells are maintained and grown within an incubator that provides the cells with a sterile, humidified, and warm environment (Figure \(\PageIndex{3}\)). The incubator is like a biosafety cabinet in that it relies upon a flow of HEPA-filtered air to prevent contamination. Air inside the incubator is filtered through a HEPA filter, warmed by the incubator and then moved through the interior of the incubator before being pulled back through the HEPA filter for filtration and recirculation. The majority of the air that moves through the incubator is taken from the environment and is a mixture of oxygen (~20%), carbon dioxide/CO2 (~0.04%), and nitrogen (~78%). The CO2 levels found in this air are inadequate for cell culture. Therefore, additional CO2 is injected into the incubator from a cylinder of compressed CO2 gas. CO2 levels for most cell cultures will range between 5% and 10%, depending on the needs of the cultured cell. Sensors inside of the incubator constantly measure the CO2 concentration and adjust the levels by injecting additional compressed CO2. Humidity levels inside of the incubator can be monitored and adjusted by the incubator. However, most incubators provide humidity to the air using a pan of sterilized water placed at the bottom of the incubator.
The air of the incubator is typically warmed to 37°C as most mammalian cells grow well at this temperature. To keep the temperature inside of an incubator stable, the incubator is “jacketed”, meaning that there is an inner chamber (where the cells are incubated) surrounded by an outer chamber (the "jacket"). The jacket prevents significant drops in temperature when the door is opened and closed. There are two kinds of jackets: water and air. The inner chamber of the air-jacketed incubator (also known as a direct heat incubator) contains air, warmed to 37°C. This type of incubator is light and cost-effective in comparison to the water-jacketed incubator. In a water-jacketed incubator, the jacket is filled with sterile water and heated to 37°C. This jacket keeps the temperature of the incubator more stable versus an air-jacketed incubator, as the water jacket will return the inner chamber's temperature to its set point much faster (i.e. within 10 minutes for water-jacketed vs. 1 hour for air-jacketed). However, the water jacket is an excellent source of contamination by micro-organisms like fungus. In addition, a water-jacketed incubator is extremely heavy and must be drained in order for the incubator to be moved.
Cell counting systems
Cell counting systems are used to determine the number of cells within a culture. The most common and inexpensive type of counting system is the hemacytometer. Originally designed and used for counting red blood cells, the hemocytometer is used for the manual counting of numerous types of cells. Refer to Lab Technique: Cell Culture for a protocol on the use of a hemocytometer. The hemocytometer consists of a thick glass microscope slide with a rectangular indentation that creates a counting chamber of 10-4 mL (Figure \(\PageIndex{4}\)). This chamber is engraved with a laser-etched grid of perpendicular lines. Cells are added to the counting chamber and visualized under the microscope. Those falling over certain parts of the counting grid are counted. With this cell count and the volume of the counting grid (i.e. 10-4 mL), the number of cells per mL can be determined. This concentration can be used to determine the total number of cells within the culture.
Automated cell counters use advanced technology to accurately and efficiently count and analyze cells isolated from a culture. Most counting systems are image-based, using cameras and microscopes to capture high-resolution images of the sample. Software processes the images to identify and count cells based on their size, shape, or staining. Other systems use measures changes in electrical resistance when the cells pass through a small aperture or channel cells through a micro-fluidic device where they are imaged, electrically measured, or optically analyzed. Automated counting systems eliminate the manual labor and potential for human error associated with traditional methods like hemacytometry. They are significantly more accurate and faster and easier to use. However, they are markedly more expensive that hemacytometers.
Inverted microscope
An inverted microscope is a type of light microscope where the optical arrangement is reversed compared to a conventional (upright) microscope. The light source, with its condenser, is located above the stage, while the objectives are positioned below the stage (Figure \(\PageIndex{5}\)). This design is ideal for observing living cells in their culture dishes, culture flasks, or multi-well plates.
Cryopreservation Systems
Cell culture facilities will often remove cells from culture in order to store them over the long-term. Storing cultures allows the facility to remove cells from culture when they are not needed and re-introduce them when they are. This saves time, materials, and money. Long-term storage of cell cultures is done at extremely low temperatures and is referred to a cryopreservation. The simplest system for cryopreservation is a liquid nitrogen tank or Dewar. The liquid nitrogen inside of the tank (approximately -320°F or -196°C) provides the optimal temperature for long-term storage of cells. Modern tanks can hold hundreds of cryopreservation samples. The samples are placed into boxes and the boxes are stacked onto racks that are lowered into the tank. The bottom of the tank holds several inches of liquid nitrogen that produces a vapor layer that is cold enough to keep the samples frozen. The liquid nitrogen inside of the tank is replenished on a regular basis.
Additional Culture Equipment
- Water bath: necessary to warm reagents used in cell culture
- Centrifuge: for the collection of culture cells
- Refrigerator (4°C) and freezer (–20°C): for storage of culture reagents
- Sterilizing equipment (i.e., autoclave)
- Aspiration pump (peristaltic or vacuum)
The equipment of a cell culture facility includes specialized equipment like the biosafety cabinet and incubator, in addition to more common types of lab equipment like water baths, microscopes, and centrifuges.
Some important concepts to remember are:
- cell culture equipment is used inside of a dedicated cell culture room where the conditions help to maintain the sterile nature of the culture
- the biosafety cabinet (BSC) creates a flow of filtered air that prevents culture contamination and protects the user from exposure biohazards
- a BSC is not a laminar flow hood
- a cell culture incubator creates and maintains the optimal temperature, humidity, and carbon dioxide levels necessary for in vitro growth of cells
- an inverted microscope allows for the visualization of live culture cells
Glossary
Air filter - a device that removes airborne particles and contaminants from the air
Biosafety cabinet (BSC) - laboratory equipment used to aseptically work with biological materials; designed to protect users from exposure to potential biohazards through its filtering and circulation of air inside of the cabinet
Contamination - the presence of unwanted microorganisms (bacteria, yeast, mold) in a cell culture or cross-contamination from other eukaryotic cell lines
Cell counter - laboratory equipment used to determine the number of cells in a sample; can be manual (i.e. hemacytometer) or automated
Cryopreservation - a process that preserves biological samples, like cultured cells, by freezing and storing them at very low temperatures
Dewar - a specialized container used for holding a cryogen, such as liquid nitrogen; cell culture Dewars are modified to store hundreds of cryo-vials containing frozen cell samples
Hemacytometer - a counting-chamber device designed and used to visualize and count cells
HEPA (high efficiency particularte air) filter - the efficiency standard of air filters that are used to remove particulate matter from the air
Incubator - laboratory equipment used to grow and maintain microbiological cultures or cell cultures under optimal temperature, humidity, and carbon dioxide levels